ALV SLE - 2000 is designed for controlled mechanical ventilation in newborn infants and children in the process of Resuscitation in the conditions of hospitals, maternity homes, research institutes child profile.
The device has the following modes for continuous ventilation:
- convection forced ventilation (CMV)
- ventilation, initiated by the patient (PTV)
- synchronized intermittent forced ventilation (SIMV)
- respiratory support with constant positive pressure (CPAP), handheld breath.
Specifications:
Respiration rate
1 - 125 or 126 - 250 breaths per minute is selected by means of a special switch
The inspiratory time
0.1 - 3.0 or 0.01 - 0.3 sec.
The ratio of I: e (inhale: exhale)
9.9: 1 x 1: 9.9 calculated from plants respiration rate and inspiratory time
Adjustable sensitivity apparatus on respiratory effort of the patient
Oxygen Blender
setting range 21 - 100% O2 with an accuracy of no more than 3%
Pressure indicator with display
Selects digital display maximum, medium or minimum pressure respiratory tract
Respiratory pressure regulator
Regulates the pressure leading to the installation of the respiratory pressure in the circuit. Range 0 - 60 cm H20
Advantages:
- Exclusive pneumatic effect
- The unique besklapannaja system reduces the problem of resistance and elasticity
- To maintain constant pressure wave shape at all speeds
- Sensitive trigger respiratory tract
- The built - in oxygen Analyzer with digital display
- High pressure emergency shut - down
- Ability to filter the exhaled gases
- Optional additional issue of mixed flow
- Optional conclusion 5 423 (compatible with 232 KB)
- Output wave form similar to the standard
- Selection of square shape wave or form a slight recovery, the selector switch
- The possibility of using other gases (e. g. , nitric oxide)
- Modular design for ease of maintenance